Advanced 2 Block Dry Block Heater with NIST Traceable Certificate, 230V

Part # 949703

  • CE mark indicates that this product conforms with the essential health and safety requirements set out in European Directives

Part # 949703

  • Exceptional temperature uniformity and stability
  • Optional external temperature probe
  • Holds interchangeable modular blocks
  • 5 year industry leading warranty

Talboys Advanced Digital Dry Block Heater is designed for applications that require repeatable results and superior temperature stability. This multi-purpose unit is ideal for incubation and activation of cultures, enzyme reactions, immunoassays, melting/boiling points and a wide variety of other laboratory procedures. Integral support rod holder with locking knob accepts optional External Temperature Probe Kit. Optional External Temperature Probe Kit monitors actual block or sample temperature.

1 Block Dry Block Heater accepts separate interchangeable modular blocks, accommodating various tube sizes from 0.2mL micro-tubes to 50mL centrifuge tubes. Each block has a thermometer well for measuring block temperature. Efficient heating due to close tube-and-block contact for maximum heat retention. Dry Block Heaters require Talboys modular heating blocks for operation. Modular blocks are sold separately. Dry Block Heaters include a 5 year, industry leading limited warranty on parts and labor.

NIST Traceable Calibration Certificate provided by Troemner's ISO/IEC 17025 accredited laboratory. Multiple data points within the temperature range and time function are reported on the certificate with the associated uncertainties.

  • up
  • down

Availability: In stock

Price: $1,374.00

Advanced 2 Block Dry Block Heater with NIST Traceable Certificate, 230VAdvanced 2 Block Dry Block Heater with NIST Traceable Certificate, 230V
Move your mouse over image or click to enlarge
Product Details
Part #949703Temperature RangeAmbient +5°C to 120°C ApplicationsDenaturings proteins, DNA applications, ELISA and other immunoassay studies.
Control TypeDigitalTemperature Stability at 37˚C± 0.2° CCertificationsCE
Block Capacity2 blockUniformity Within the Block at 37˚C± 0.2° CWarranty5 year, industry leading limited warranty on parts and labor.
Frequency50/60 HzUniformity Across Similar Blocks at 37˚C± 0.1° CUnit Supplied WithUnits include a 92" (234cm) detachable cord with Euro (type) plug. NIST Traceable Calibration Certificate provided by Troemner's ISO/IEC 17025 accredited laboratory. Multiple data points within the temperature range and time function are reported on the certificate with the associated uncertainties.
Voltage230VHeat-Up Time to 100˚C50 min.OriginManufactured in the USA
Amps0.9Dimensions15.4 x 8 x 3.5"
(39.1 x 20.3 x 8.9 cm)
Price

Price: $1,374.00

Watts210Operating ConditionsUnits can be run in conditions from 18 to 33°C (64 to 91°F), 20% to 80% relative humidity, non-condensing  


Operating & Safety Features
Microprocessor ControlPID temperature control, with optional RTD probe, offers a temperature stability as low as ± 0.1°C with temperature uniformity as low as ± 0.1° C. Samples are heated to temperature quickly and accurately. Temperature is adjusted in ± 0.1°C increments.
LED DisplayTouch pad controls with easy-to-read, independent LED displays for temperature and time. Provides repeatable and accurate results every time and is easily visible across lab benches. Shows set-point and actual temperature. Timer will display elapsed time or, when programmed to user defined limit, will shut off unit when time reaches zero. Display will show last used settings, even after power has been turned off.
Audible AlarmIn timed mode, alarm will sound when time reaches zero and when unit reaches set-point temperature.
Temperature Calibration ModeAllows user to calibrate unit to an external temperature device.
Caution Hot IndicatorHot warning symbol light is illuminated when temperature is above 40°C.
Overshoot ProtectionIf the unit exceeds the set temperature by 10°C the unit will automatically stop heating.
CAUTION!To avoid possible electrical hazard, do not fill well or block opening with water or other fluids. Unit is designed as a dry bath/incubator.


Voltage Explanation

For use in the United States, typical voltage is 120V. 120V units are supplied with standard 3-prong grounded plug.


For use outside the United States or for 230V applications, 230V units are supplied with a Euro type plug. If you require a different plug type, please specify the part number listed below in the comments when ordering. For plug types not displayed below, please contact Inside Sales at +1.856.686.1600 or via email at troemner_web@troemner.com. All plugs supplied are standard IEC 320 type connector.

Click here for complete voltage details and plug types by country

Literature & Reference